Needle flow tester:
International standard: ISO 7864:2016 《Sterile hypodermic needles for singleuse - Requirements and test methods》Aiguilles hypodermiques steriles, non reutilisables - Exigences etmethodes d'essai。
Test Item: Determination of flow rate through the needle
Control System: PLC
Human-Machine Interface Platform: ARM Architecture
Operating System: Linux
Pressure Output Range Above local atmospheric pressure, 0kPa~200kPa, error not greater than ±2.5% of the reading; Resolution: 0.1kPaDisplay Method: LCD digital display; the pressure output value can be arbitrarily set, and the verification error is not greater than ±2.5% of the reading.Test Duration: 1 second to 99.9 minutes, error not greater than ±1s
For catheters with a nominal outer diameter of less than1.0 mm, the flow rate should be 80% to 125% of the manufacturer's marked value;
For catheters with a outer diameter greater than or equal to 1.0 mm, the flow rate should be 90% to 115% of the manufacturer's value.
Sensor: High-precision force sensor (with software and hardware protection functions)
Fixture: High-quality 304 stainless steel
Force Measurement: Built-in
Printer: On-board dot matrix printer
Data Storage: Can be imported to a computer via USB flash drive for storage
Total Power:1000W
Power Supply: AC 90V-240V/50Hz
Net Weight: 11kg
External: 400×290×240 (mm)