Selectable modes  | Rate mode,  Time mode ,Body weight mode,  | 
 KVO rate    | 0.1ml/h-5.0ml/h, adjustable, choose “OFF”to cancel     | 
Injection accuracy   |  ±2%    | 
Bolus rate    |  5ml: 100ml/h        20ml: 400ml/h  10ml: 200ml/h     30ml: 400-600ml/h    50ml:400-1500ml/h    | 
Injection rate range  | 5ml: 0.1ml/h---100ml/h         10ml: 0.1ml/h---200ml/h 20ml: 0.1ml/h---400ml/h        30ml: 0.1ml/h---600ml/h 50(60)ml: 0.1ml/h---1500ml/h   | 
Drug names    | More than 20 drug names, such as Penicillin, Vancomycin, Pitocin , Heparin, Atropin, Lidocaine, Dopamine, Epinephrine,Insulin, etc.   | 
Applicable Syringes  | All brands  5, 10, 20, 30 and 50 (60) ml   | 
| Injection increment | 0.1ml/h    | 
Body weight mode  | Body weight: 0.1~300kg    Drug limit: 0.1~999.9mg Volume limit: 0.1~999.9ml   Dose:0.1~9999.9mg (Subject to unit, dosage, liquid, weight and syringe) Unit: mg/kg/h, ug/kg/min  | 
Rate mode   | 0.1~1500ml/h(max value depends on the syringe size)   | 
| Time mode | Volume limit: 0.1-999.9ml Time: 1-2000min | 
Accumulated volume   | 0.1~9999.9ml    | 
Preset range   | 0.1~9999.9ml    | 
Occlusion value   | 0.04Mpa – 0.16Mpa   | 
Preset time  | 0-99hour, 59 minutes   | 
Alarm function    | NEAR, OVER, OCCL.,, no AC power supply, low battery, serious low battery, error 1/2/3, preset volume alarm, syringe disengaged  | 
Panel instructions  | Infusion rate ,  Syringe pump , Battery capacity , Battery charging indication , AC power indicator ,Accumulated injection volume, BED NO. , Alarm information, etc  | 
Battery    | Rechargeable lithium polymer battery 7.4V, 3000mA, working for 4 hours(25ml/h) after being fully charged    | 
Power supply  | AC100~240V   50/60HZ   | 
Max power consumption     | 25VA   | 
Classification    | Class II, type BF   | 
EMC    | Meet the requirements of IEC60601-1-2(2007 version)     | 
Electrical safety   | Meet the requirements of EN60601-1:1990+A1:1993+A2:1995   | 
Storage condition  |  Environment temperature: -20℃~+55℃, air pressure (50~106)kPa,  relative humidity≦95%  | 
Operation condition   |  Environment temperature: 5℃~+40℃, air pressure (86~106) kPa,  relative humidity≦80%  | 
Gross weight   | <2.8kg   | 
Overall maximum size   | 300L*139W*100mmH   | 