Welcome!


Infusion set leak tester:
International standard: ISO 8536-4:2019 《Infusion equipment for medical use —
Part 4: Infusion sets for single use, gravity feed》
Test item: A.3 Test for leakage
Control system: PLC
HMI platform: ARM architecture
Operating system: Linux
Display mode: Pressure output value can be arbitrarily set, verification error not greater than2.5% of the reading.
Positive and negative pressure output range: -94kPa~200kPa (adjustable), LCD digital display, error greater than ±2%.
Positive pressure retention of the measurement system: Cannot drop by more than 1kPa within 1 minute.
Test duration: second to 99.9 minutes, error not greater than ±1S.
Sensor: High-precision force sensor (with software and hardware protection functions)Fixture: High-quality 304 stainless steel
Force measurement mechanism: Built-in
Printer: Onboard dot matrix printer
Data storage: Can be to a computer via USB flash drive for storage
Total power: 1000W
Power supply: AC90V-240V/0Hz
Net weight: 15kg
Dimensions: 400×290×275 (mm)
Contact Us